Product details
Overview
MPX2301DT1 from Freescale Semiconductor is a miniature, differential pressure sensor with 0–40 kPa full-scale range, integrated thin-film temperature compensation and calibration, ratiometric output, and ISO 10993-compliant polysulfone housing. It delivers 5.0 μV/V/mmHg sensitivity, ±1.5% VFSS linearity + hysteresis, and 1.0 ms response time for invasive medical pressure monitoring applications.
For engineers reviewing the MPX2301DT1 datasheet, MPX2301DT1 pinout, MPX2301DT1 application, or MPX2301DT1 equivalent, key selection criteria include biocompatible packaging, defibrillation-safe gel isolation, ratiometric supply dependency (6 V ±0.2 V), low zero-offset drift (±9.0 μV/°C), and Chip Pak package compatibility with disposable medical sub-modules.
Technical Context
The MPX2301DT1 employs piezoresistive silicon sensing element with backside pressure input and front-side exposed die/wire bonds requiring customer-provided mechanical protection. Its on-chip thin-film compensation circuitry corrects for temperature-induced sensitivity and offset drift across +15°C to +40°C operating range.
Output is fully differential (VOUT+, VOUT–) with 270–330 Ω output impedance and 1800–4500 Ω input impedance, referenced to a 6 V regulated supply. The device exhibits ratiometric behavior - output voltage scales linearly with supply voltage - and supports RCAL-based offset calibration using a 150 kΩ resistor.
Key Specifications
| Parameter | Value and Actual Design Meaning |
|---|---|
| Pressure Range | 0–40 kPa (0–300 mmHg) differential; defines usable clinical pressure window for blood pressure and infusion monitoring. |
| Supply Voltage | 6.0 V ±0.2 V regulated; ratiometric output scaling requires tight regulation to avoid measurement error. |
| Sensitivity | 5.0 μV/V/mmHg typical; determines signal amplitude per unit pressure at given excitation voltage. |
| Linearity + Hysteresis | ±1.5 %VFSS max; specifies maximum deviation from ideal straight-line output over full scale at 101–200 mmHg. |
| Response Time | 1.0 ms (10% to 90%); enables real-time capture of dynamic pressure waveforms in arterial monitoring. |
| Operating Temperature | +15°C to +40°C; validated for human-body-adjacent use in clinical environments, not extended industrial range. |
| Zero Offset Drift | ±9.0 μV/°C; quantifies thermal-induced baseline shift requiring system-level compensation or calibration. |
Pinout & Package
MPX2301DT1 uses the Chip Pak package (Case 98ASB13355C), a surface-mount, biocompatible module with 4-terminal configuration and backside pressure port. Dimensions: 6.10–6.60 mm × 8.89–9.40 mm × 3.56–3.81 mm (L×W×H).
| Pin/Terminal | Circuit Role | Design Meaning |
|---|---|---|
| 1 - VS | Voltage supply input | Accepts regulated 6 V ±0.2 V; powers internal sensing element and compensation circuitry. |
| 2 - VOUT+ | Differential output positive | Active high-side output; used with VOUT– to reject common-mode noise in medical cables. |
| 3 - VOUT– | Differential output negative | Active low-side output; paired with VOUT+ to deliver true differential analog signal. |
| 4 - GND | Analog ground reference | Return path for supply and output circuits; must be isolated from noisy digital grounds in PCB layout. |
Key Features
| Feature | Design Value |
|---|---|
| On-chip temperature compensation | Thin-film resistors correct sensitivity (±0.1%/°C) and full-scale span (±0.1%/°C) without external circuitry. |
| Ratiometric output | Output voltage scales directly with VS; eliminates need for precision voltage references in ADC front-end design. |
| Biocompatible construction | Polysulfone housing (ISO 10993-5/10/11) and USP Class V silicone gel enable ethylene oxide sterilization and saline exposure. |
| Differential output architecture | VOUT+/VOUT– pair provides inherent common-mode rejection (>60 dB typical), critical for EMI-prone hospital environments. |
| Chip Pak modular form factor | Surface-mountable, tape-and-reel packaged sub-module designed for integration into disposable catheter assemblies or transducer housings. |
Applications
| Infusion Pump Pressure Monitoring | Blood Pressure Transducer Assembly |
|---|---|
|
Use Scenario: Real-time detection of occlusion, air-in-line, or free-flow conditions during IV therapy. IC Role / Device Role / Timing Role: Differential pressure sensor measuring pressure drop across flow restrictor or tubing segment. Use Value: 1.0 ms response time captures rapid pressure transients; ±1.5% accuracy ensures reliable alarm thresholds per IEC 60601-2-24. |
Use Scenario: Disposable, single-patient blood pressure transducer connected to arterial line. IC Role / Device Role / Timing Role: Primary pressure-sensing element converting fluid column pressure to calibrated analog output. Use Value: Biocompatible gel and polysulfone housing meet AAMI BP-22 defibrillation safety requirements and ISO 10993 biological testing. |
| Pressure Catheter Tip Sensing | Patient Monitoring System Sub-Module |
|
Use Scenario: Miniaturized pressure sensing at distal tip of diagnostic catheters for intracardiac or intravascular measurements. IC Role / Device Role / Timing Role: Backside-pressurized sensing die embedded in catheter tip assembly with front-side die protection. Use Value: Chip Pak's 6.6 mm × 9.4 mm footprint fits within 8 Fr catheter profiles; low power (1.0 mA) extends battery life in portable monitors. |
Use Scenario: OEM integration into multi-parameter patient monitors for non-invasive or invasive pressure channels. IC Role / Device Role / Timing Role: Calibrated analog front-end component providing ratiometric output to precision ADC subsystem. Use Value: RCAL-based offset calibration simplifies production test; tape-and-reel packaging supports automated SMT assembly. |
Equivalent & Alternatives
The following parts are listed as comparable options for similar differential pressure sensing applications.
| Alternative Part | Technical Difference | Application Difference | Selection Advice |
|---|---|---|---|
| NXP MPXV5004DP | 0–100 kPa range, 5 V supply, single-ended output, no integrated gel isolation; TO-92 package. | Lacks biocompatible gel and ISO 10993 housing; unsuitable for direct fluid contact or sterilization. | Select when higher pressure range and cost-sensitive industrial use outweigh medical compliance needs. |
| TE Connectivity MS5837-02BA | I²C digital output, 0–2 bar range, stainless steel housing, 0.2% FS accuracy, 24-bit resolution. | Requires microcontroller interface and firmware; no analog ratiometric output or medical-grade gel encapsulation. | Select when digital interface, higher accuracy, and corrosion resistance are prioritized over analog simplicity and disposability. |
Compared with MPX2301DT1, the MPXV5004DP offers wider pressure range but lacks medical qualification and differential output, while the MS5837-02BA provides digital precision and ruggedness but removes analog ratiometric operation and biocompatible integration - making MPX2301DT1 uniquely suited for disposable, sterilizable, analog-front-end medical transducers.
Availability
MPX2301DT1 is available at Aetrix Electronics and suitable for infusion pump pressure monitoring, blood pressure transducer assembly, and pressure catheter tip sensing requiring stable component supply, traceable sourcing, and lifecycle continuity for FDA-regulated medical device manufacturing.
Supply support for MPX2301DT1 includes scheduled delivery planning, volume procurement assistance, BOM continuity management, traceable sourcing, and lifecycle availability coordination for OEM customers, industrial embedded developers, connected-device designers, and electronics production programs.
Manufacturer
Freescale Semiconductor (now part of NXP Semiconductors) is a fabless semiconductor company specializing in embedded processing, sensors, and automotive ICs, with deep expertise in medical and industrial analog signal conditioning.
The MPX2301DT1 belongs to Freescale's MPX2300 series of compensated pressure sensors, engineered specifically for disposable, biocompatible, invasive medical applications requiring regulatory compliance, sterilizability, and high signal integrity in noisy clinical environments.
FAQ
What is the pressure port orientation and mechanical interface requirement for MPX2301DT1?
The MPX2301DT1 applies pressure to the backside of the Chip Pak package - the side opposite the printed markings and pins. The front side contains exposed die and wire bonds that must be protected by the customer's housing. A compliant medical-grade gel layer transmits pressure uniformly to the diaphragm while electrically isolating connections from saline or other fluids. Proper sealing against the backside port is essential to prevent leakage and ensure accurate differential measurement in MPX2301DT1 installations.
Does MPX2301DT1 require external calibration components?
No, the MPX2301DT1 integrates thin-film temperature compensation and factory calibration, eliminating the need for external trim pots or calibration ICs. However, RCAL-based offset verification is supported using a 150 kΩ resistor between VS and VOUT+ to confirm zero-pressure output alignment. This optional check aids production test but does not replace the internal calibration - the MPX2301DT1 delivers calibrated output directly from the pins under specified supply and temperature conditions.
Can MPX2301DT1 be sterilized, and which methods are validated?
Yes, MPX2301DT1 is validated for ethylene oxide (EtO) sterilization, consistent with its ISO 10993-compliant polysulfone housing and USP Class V silicone gel. Autoclaving and gamma irradiation are not recommended, as they may degrade the gel or housing material. The device's biocompatibility testing covers ISO 10993-5 (cytotoxicity), -10 (irritation/sensitization), and -11 (hemolysis), confirming suitability for short-term invasive use in MPX2301DT1-based transducers.
What is the meaning of "ratiometric" in MPX2301DT1 specifications?
"Ratiometric" means the MPX2301DT1's output voltage scales proportionally with its supply voltage (VS). For example, at 6.0 V supply, full-scale output is ~3.0 mV; at 5.9 V, it drops to ~2.95 mV. This behavior allows the same reference voltage used for VS to also serve as the ADC reference, canceling supply-induced errors. Designers must regulate VS tightly (±0.2 V) because unregulated supplies introduce direct gain error into the MPX2301DT1 measurement chain.
Is MPX2301DT1 pin-compatible with other devices in the MPX2300 family?
Yes, MPX2301DT1 shares identical pinout (VS, VOUT+, VOUT–, GND), package (Case 98ASB13355C), and footprint with MPX2300DT1 and MPX2302DT1. Differences lie solely in pressure range (MPX2300DT1: 0–20 kPa; MPX2301DT1: 0–40 kPa; MPX2302DT1: 0–100 kPa) and corresponding sensitivity/offset calibration. This allows hardware reuse across variants - a single PCB layout can support multiple MPX2300-series sensors by selecting the appropriate MPX2301DT1 variant during assembly.
MPX2301DT1 Specifications
- Product attributes
- Attribute value
- Manufacturer:
- NXP Semiconductors
- Series:
- MPX2301
- Package/Case:
- 4-SSIP Module
- Packaging:
- Tape & Reel (TR)
- Product Status:
- Obsolete
- Applications:
- Board Mount
- Pressure Type:
- Differential
- Operating Pressure:
- 5.8PSI (40kPa)
- Output Type:
- Wheatstone Bridge
- Output:
- 0 mV ~ 3.006 mV (6V)
- Accuracy:
- ±1.5%
- Voltage - Supply:
- 6V ~ 10V
- Port Size:
- -
- Port Style:
- No Port
- Features:
- Temperature Compensated
- Termination Style:
- PC Pin
- Maximum Pressure:
- 125PSI (861.84kPa)
- Operating Temperature:
- 15°C ~ 40°C
- Grade:
- -
- Qualification:
- -
- Mounting Type:
- Through Hole
- Supplier Device Package:
- -
MPX2301DT1 FAQ
1.How can I place an order for MPX2301DT1 through Aetrix?
Please submit a Request for Quotation (RFQ) for MPX2301DT1 on Aetrix. Our sales agent will provide a competitive quotation and guide you through the order confirmation once you accept the terms.
2.Are the price and stock information for MPX2301DT1 reliable?
The price and inventory of MPX2301DT1 are updated periodically and may fluctuate due to market conditions. Stock and pricing data are typically refreshed every 24 hours. Quotation validity for MPX2301DT1 is usually 5 days.
3.What payment methods are accepted for MPX2301DT1?
We accept Wire Transfer, PayPal, Credit Card, Western Union, MoneyGram, and Escrow for MPX2301DT1 transactions.
Note: Certain payment methods may incur a processing fee.
4.How is shipping managed for MPX2301DT1?
MPX2301DT1 orders can be shipped via leading logistics carriers, including DHL, UPS, FedEx, TNT, or Registered Mail.
Once your MPX2301DT1 order is processed, you will receive an email with the shipment details and tracking number.
Note: Tracking information may take up to 24 hours to appear. Express delivery typically takes 3–5 business days.
5.How can I obtain technical support or documentation for MPX2301DT1?
For technical support, including MPX2301DT1 datasheets, pinout diagrams, or application guidance, please contact our engineering support team. They can provide detailed documentation and assistance for your MPX2301DT1 requirements.
6.How does Aetrix verify that MPX2301DT1 is sourced from the original manufacturer or authorized distributors?
All MPX2301DT1 products on Aetrix are procured from qualified distributors and authorized channels. Our dedicated quality assurance team conducts strict verification, including traceability checks and, if necessary, third-party testing. This ensures that MPX2301DT1 meets industry standards.
7.What is the process for return or replacement of MPX2301DT1?
All MPX2301DT1 units undergo pre-shipment inspection (PSI). If there is an issue with MPX2301DT1, returns or replacements are accepted under the following conditions:
1.Quantity discrepancies, incorrect items, or visible external defects (such as breakage or corrosion), acknowledged by Aetrix.
2.The issue is reported within 90 days of delivery.
3.The MPX2301DT1 part is unused and in its original packaging.
Return procedure for MPX2301DT1:
1.Submit a request within 90 days.
2.Obtain a Return Material Authorization (RMA) from Aetrix.
MPX2301DT1 Tags

-
DPS368XTSA1
Infineon Technologies

-
DPS310XTSA1
Infineon Technologies

-
LPS22HHTR
STMicroelectronics

-
ICP-20100
TDK InvenSense

-
ICP-10110
TDK InvenSense

-
LPS22DFTR
STMicroelectronics

-
LPS22HBTR
STMicroelectronics

-
BMP390
Bosch Sensortec

-
BMP581
Bosch Sensortec

-
BMP384
Bosch Sensortec

-
2511020213301
Würth Elektronik

-
ILPS22QSTR
STMicroelectronics
Tech Hub
A practical engineering guide to 3.3V and 5V logic compatibility, input thresholds, resistor dividers, translator ICs, MOSFET level shifting, I2C pull-ups, timing limits and power-sequencing risks.
The 74HC595 uses push-pull logic outputs, while the TPIC6B595 uses 50 V open-drain DMOS sinks for higher-power loads. This guide compares timing, current limits, 3.3 V interfacing, load wiring, thermal…
The 74HC595 converts serial data into eight stable parallel outputs. This guide covers pin functions, shift and storage timing, OE and MR behavior, drive-current limits, cascading, voltage compatibilit…
A technical comparison of level-sensitive latches and edge-triggered flip-flops, covering timing windows, setup and hold limits, master–slave operation, time borrowing, race-through, HDL inference and…
A D latch stores one bit while Enable controls when data can pass. This reference covers gate-level operation, truth tables, transparency, setup and hold timing, LE versus OE, common ICs and practical …
An SR latch stores one bit through cross-coupled feedback. This engineering reference covers NOR and NAND implementations, truth tables, forbidden-state recovery, gated operation, switch debouncing, fa…
Latch circuits retain one bit through feedback. This technical reference covers SR and D latches, truth tables, transparency, timing limits, latch-versus-flip-flop behavior, applications and common log…
An engineering guide to LED driver operation, constant-current and constant-voltage outputs, linear and switching topologies, dimming, IC selection, calculations, replacement compatibility, and fault c…
Operational amplifier guide covering op amp basics, feedback, ideal vs real op amps, common configurations, buffer circuits, offset, bias current, gain-bandwidth, slew rate, rail-to-rail limits and sel…
Jumper cables guide covering safe connection order, red and black clamp placement, final ground connection, cable gauge, length, clamp quality, copper vs CCA cables, jump starter comparison and battery…